DC-STIMULATOR MC*
Programmable, Multi-Channel Direct, Alternating and Random Noise Current Stimulator
Focus of application is in research into multi-channel, controlled, supervised and stimulation
Transcranial Electrical Stimulation (TES) using weak electric direct, alternating and random noise currents (tDCS/tACS/tRNS) over a period of several minutes changes the electrical charge in the nerve cell membranes and has a direct influence on neurotransmitter channels. This serves to strengthen or diminish the excitability of the brain cells.
The multi-channel DC-Stimulator MC allows laptop-controlled, full-band stimulation from independent electrical sources using any desired signal type in the range of 0-1,000 Hz and currents of between 50-5,000 µA with a freely adjustable phase.
The DC-Stimulator MC can also be used during functional magnetic resonance imaging (fMRI) and, in addition, can be combined with the NEURO PRAX® TMS/tDCS allowing full-band DC-EEG to be recorded during multi-channel transcranial electrical stimulation.
- Up to 16-channel, bipolar (AC) stimulation
- For transcranial direct current stimulation (tDCS), transcranial alternating current stimulation (tACS), cranial electrical stimulation (CES), galvanic vestibular stimulation (GVS) and transcranial random noise stimulation (tRNS)
- 4 channels, capable of alternating current, bipolar stimulation possible (8, 16 channels)
- Panel PC for the use and programming of stimulation modes and stimulation sequences
- Various types of stimulation can be selected and combined, continuous stimulation, cyclical switching on and off of stimulation, sinusoidal stimulation (up to 1,000 Hz)
- Import of any stimulation sequences into the software to control the DC-Stimulator MC with customer-specific signal sequences
- External trigger input and output (optional)
- MRI extension (optional)
DC-STIMULATOR MC* Specifications
- Currents and wave forms of up to ± 5,000 µA
- AC current and adjustable up to 3,000 µA (peak-to-peak)
- Frequencies of up to 1,000 Hz selectable and phase freely adjustable
- Freely selectable application duration*
- 16-bit D/A conversion
- Time resolution < 1 ms (sample rate 24,000 sps)
- Recording of stimulation sequences with 8,000 measurements/second with max. 1% relative direct current fault tolerance.
- Max. 0.02% direct current fluctuation
- Current variance during stimulation < 0.02%
- Voltage limit of 26V
- Dimensions: 16.5"W x 15.6"D x 6.7"H
- Weight: 9.3 lbs
DC-STIMULATOR MC* Brochure
DC-STIMULATOR MC* Brochure (Coming Soon)