DC-STIMULATOR MR*
Programmable Direct and Alternating Current Stimulator
Conducting non-invasive transcranial direct- and alternating current stimulation (tDCS, tACS) specifically during magnetic resonance imaging (fMRI)
The DC-STIMULATOR MR meets the highest safety standards thanks to (hardware- and software-based) multistage monitoring of the current path. By continuously monitoring electrode impedance it can detect insufficient contact with the skin and automatically terminate stimulation.
Scientific studies conducted in recent years have shown transcranial direct current stimulation (tDCS) to be a recognized method of therapy for various disorders of the brain.
Alternating current and random noise current stimulation (tACS, tRNS) are currently in the research and validation phases. Transcranial electrical stimulation (tDCS, tACS and tRNS) can also be performed during functional magnetic resonance imaging to localize the exact position of cortical activation.
- 1 channel, unipolar (DC) and bipolar (AC) stimulation
- Use during MRI, no interference of the MRI images during EPI sequence
- 4 standard modes - single (continuous stimulation) - pulse (cyclical stimulation activation/deactivation) - sinus (sinus wave) - noise (normally distributed)
- "Study mode" for blind processing of genuine and 'pseudo' stimulation (optional)
- Adjustable current (DC) up to 4,500 µA in increments of 25 µA
- Adjustable current (AC) up to 3,000 µA (peak-to-peak)
- Additional MRI protection resistor of approx. 5 kOhm in each electrode
- External trigger input and output (optional)
DC-STIMULATOR MR* Specifications
- Internal 12bit D/A conversion
- Internal time resolution 1 ms (sample rate 1,280 sps)
- "Single" stimulation mode: duration 15-1,800 s, increment 15 s, duration of fade in/fade out 1-120 s, increment 1 s
- "Pulse" stimulation mode: duration of complete pulse cycle/interstimulus interval (ISI) 300-2,000 ms, increment 100 ms, pulse width 200- (ISI-100), increment 100 s, number of pulse ccles 1-500
- "Sinus" stimulation mode: adjustable current up to 3,000 µA (p-p) in 25 µA increments, offset of 0 to ± 1,000 µA, increment 10 µA, frequencies up to 250 Hz, increment 0.01 Hz, adjustable phase 0-360º in 5º steps, application time adjustable up to 30 min
- "Noise", "noise LF", "noise HF" stimulation mode: adjustable current up to ± 1,500 µA (p-p), offset of 0 up to ± 1.000 µA, increment 50 µA, duration 0-1,800 s in 5 s increments, current adjustable over period of 0-120 s to reach and leave oscillation level
- Max. voltage limitation ± 20V
- Dimensions: 5.3"W x 8.9"D x 2.2"H
- Weight (incl. batteries): 1.8 lbs
DC-STIMULATOR MR* Brochure
DC-STIMULATOR MR* Brochure (Coming Soon)