The PowerMAG LAB 100* is capable of emitting single pulses, repetitive TMS as well as Theta Burst protocols. Pulse timing and intensity is always precise and allows for reproducibility. The stimulator can be controlled by different means and is easily integrated into your existing setup.
Theta Burst with up to 100 Hz possible
no decrease of pulse intensity during stimulation
exact timing
Maximal 30 Hz
100% intensity up to 15 Hz
Full external controllability
Half wave and full wave available
Current direction in the coil invertible
Technical Specifications | PowerMAG LAB 100* |
---|---|
Maximal frequency | 100 Hz |
Frequency controllability gradation | 0.1 Hz |
Maximal frequency @ 100% intensity | 15 Hz |
Maximal intensity | 100% (15 Hz), 70% (25 Hz) |
STIMULATION PARAMETERS | |
Magnetic induction | up to 4 tesla (depending on coil) |
Maximum pulse power | 160 Joule |
Pulse shapes | Half & full sine-wave |
Current direction in the coil | Invertible on the equipment |
Rest time of the stimulation pulse | 80 µs (PA) |
Pulse duration | 160 µs |
INPUT/OUTPUT INTERFACE | |
Trigger input | BNC-connector (TTL pulse, active high) |
Trigger output | BNC-connector (TTL pulse, active high) |
External Analogue Control | Via BNC & Control Splitter Box (Trigger IN, Trigger OUT, Intensity, Pulse Shape) |
External Serial Control | Via USB & PC-Interface (Matlab, Python, Presentation, ...) |
QUALITY "MADE IN GERMANY" | |
Number of discharges @ 100% intensity | 2,000,000 |
Electrical safety | Protection class 1, Type BF |
Hardware classification to MDD 93/42/EEC | Class IIa |
Hardware certification | CE 0494 |
OPERATING PARAMETERS | |
Power supply voltage | 100/110/120/220/230 V~, 50/60 Hz (PowerUnit possibly necessary) |
Power consumption | 50 VA in standby mode, 800 VA nominal rating, 2000 VA |
Operating Temperature | 0 °C to +40 °C |
Dimensions | 490 x 260 x 600 mm (width, height, depth) |
Weight | 38 kg |
Items marked with* are investigational devices and for research use only. CAUTION - Investigational Device. Limited by Federal (or United States) law to investigational use.